Chapter 5
Characterization of Molecular Weight
5.1 Introduction
With the exception of a few naturally occurring polymers (such as proteins and DNA), all polymers consist of molecules with a distribution of chain lengths. Compared to molecules such as water that have a definite structure (H2O) and molecular weight (18), the polymer chains making up a sample consist of molecules with a range of molecular weights. For example, poly(ethylene glycol) or PEG is often used at a range of molecular weights, around 400 (which is actually just an oligomer of approximately nine repeat units), PEG is a viscous liquid, while with increasing molecular weight, PEG becomes waxy (around 600), and can be formed into a solid powder (around 3000). Because the molecular weight is strongly related to the mechanical and chemical properties of polymers, it is necessary to characterize the entire distribution quantitatively, or at least to define and measure average chain lengths or molecular weights for these materials. Extensive reviews are available [1, 2] concerning the effects of molecular weight and its distribution on the mechanical properties of polymers.
First, the molecular weight of a single polymer chain, M, is easily calculated, if the chain length, x, and chemical composition of the repeat unit (and thus the molecular weight of the repeat unit, Mr) are known:
Note that the terms chain length and degree of polymerization can be used interchangeably, and both are represented by x.
When considering a collection of polymer chains, an average molecular weight must be considered. However, the method for determining this average depends on how you count. In normal averaging (e.g., your grades in a class), you sum up each of the individual chain molecular weights (or scores) and divide by the number of chains (or assignments):
(5.2)
This is a perfectly valid way to calculate molecular weight, but we must consider statistics and distribution patterns to get a more accurate picture of how polymers behave, so this chapter will discuss different ways of calculating molecular weights, introduce distribution profiles, and discuss laboratory methods used to measure or estimate the average molecular weight of polymer samples.
5.2 Average Molecular Weights
Since nearly every collection of polymer molecules contains a distribution of molecular weights, a number-average molecular weight may be defined in an analogous fashion to Equation (5.3):
Although this equation appears complicated, if you work with it, you will see that it reduces to the common equation used to average grades, except that it gives the average length of all the chains in a polymer sample. Any analytical technique that determines the number of moles present in a sample of known weight, regardless of their size, will give the number-average molecular weight.
Rather than count the number of molecules of each size present in a sample, it is possible to define an average in terms of the weights of molecules present at each size level. This is the weight-average molecular weight, . places an emphasis on the larger molecular weight chains, which is particularly useful when considering the mechanical properties of polymers, as the larger chains contribute more to a polymer's strength than smaller-sized chains.
Consider a bowl of fruit that contains three apples, five oranges, and a bunch of grapes. If you consider each grape as a separate fruit, the average size (“number average molecular weight”) will be much closer to the size of a grape, but the majority of the mass in the bowl is due to the larger apples and oranges, so a weight-average size will represent that the larger fruits make up more of the mass. For a polymer, the weight-average molecular weight, , is
Analytical procedures that, in effect, determine the weight of molecules at a given size level result in the weight-average molecular weight. Thus, is greater than and would be closer to the size of the apples and oranges.
The number-average molecular weight is the first moment of the molecular weight distribution, analogous to the center of gravity (the first moment of the mass distribution) in mechanics. The weight-average molecular weight, the second moment of the distribution, corresponds to the radius of gyration in mechanics. Higher moments, for example, , the third moment, may also be defined and are used occasionally.
It is sometimes more convenient to represent the size of polymer molecules in terms of the degree of polymerization or chain length x, rather than molecular weight, as related by Equation (5.1). Thus, and can also be used to represent average polymer sizes, and
where m = molecular weight of a repeating unit (also sometimes called Mr)
= number-average degree of polymerization or chain length
= weight-average degree of polymerization or chain length
(These relations neglect the difference between the end groups on the molecule and the repeating units. This is perfectly justifiable in most cases since the end groups are an insignificant portion of a typical large polymer molecule.) It is also possible to use these relations for copolymers, but m should be modified to reflect the average composition of a repeat unit in the chain. In terms of chain lengths, Equations (5.3) and (5.4) become
It may be shown that ≥ ( ≥ ). The averages are equal only for a monodisperse polymer, where all of the chains are exactly the same length. The ratio / = / is known as the polydispersity index, PI. PI is a measure of the breadth of the molecular weight distribution. A monodisperse sample would have a PI of 1.0, but typical values range from about 1.02 for carefully fractionated polymers synthesized by anionic addition reactions to over 50 for some commercial polymers.
5.3 Determination of Average Molecular Weights
Everything would be quite easy if we knew the molecular weight of each chain in a polymer sample, but this is quite difficult to determine. This challenge has led to the development of several different techniques that attempt to measure or at least estimate the molecular weight of polymers [3]. This section covers some of the more common ways to determine molecular weight, including an overview of the operation of the instruments and (hopefully) giving an appreciation for the associated advantages and limitations.
In general, techniques for determination of average molecular weights fall into two categories: absolute and relative. In the former, measured quantities are theoretically related to the average molecular weight; in the latter, a quantity is measured that is in some way related to the molecular weight, but the exact relation must be established by calibration with one of the absolute methods.
Another important way to categorize these techniques is the type by which the average molecular weight is determined: or (or a viscosity-average molecular weight, ). Another type of molecular weight, , molecular weight between crosslinks, is only applicable for polymer networks and is discussed in later chapters. The techniques covered in this chapter are listed in Table 5.1, along with the type(s) of molecular weight determined by each.
Method | Type of Molecular Weight Determined |
End group analysis | |
Boiling-point elevation | |
Freezing-point depression | |
Osmotic pressure | |
Light scattering | |
Ultracentrifugation (sedimentation) | |
Viscometry | |
Gel permeation chromatography | ,
, MWD, and PI |
5.3.1 Absolute Methods
These methods allow the determination of the total number of moles of chains in a sample, N. If the total sample weight, W, is known,
(5.8)
5.3.1.1 End-Group Analysis
If the chemical nature of the end groups on the polymer chains is known (as in many condensation polymers), standard analytical techniques may sometimes be employed to determine the concentration of the end groups and thereby of the polymer molecules, giving directly the number-average molecular weight. For example, in linear polyesters formed from a stoichiometrically equivalent batch of monomers, there are, on average, one unreacted acid group and one unreacted –OH group per molecule. These groups may sometimes be analyzed by appropriate titration. If an addition polymer is known to terminate by disproportionation (see Chapter 9), there will be one double bond for every two polymer molecules, which may be detectable quantitatively by halogenation or by infrared measurements. Other possibilities include the use of a radioactively tagged initiator that remains attached to the end of each chain.
These methods have one drawback in addition to the necessity of knowing the nature of the end groups. As the molecular weight increases, the concentration of end groups (number per unit volume) decreases, and the measurement sensitivity drops off rapidly. For this reason, these methods are generally limited to smaller polymers in the range of <10,000.
5.3.1.2 Colligative Property Measurements
You may recall from an introductory chemistry course that addition of salts to a liquid can lead to boiling-point elevation or freezing-point depression. When a solute is added to a solvent, it causes a change in the activity and chemical potential (partial molal Gibbs free energy) of the solvent. The magnitude of the change is directly related to the solute concentration. For example, when pure water is boiled, the chemical potentials of the liquid and the vapor in equilibrium with it are the same. If, now, some salt is added to the water, it lowers the chemical potential of the liquid water. To reestablish equilibrium with the pure water vapor above the salt solution, the temperature of the system must be raised, causing a boiling-point elevation. In a similar fashion, the addition of ethylene glycol antifreeze depresses the freezing point of water.
Freezing-point depression, boiling-point elevation, and, a third technique, osmotic pressure, may be used to determine the number of moles of polymer per unit volume of solution and thereby establish the number-average molecular weight. The following are the relevant thermodynamic equations for the three techniques:
(5.9)
(5.10)
where
c = solute (polymer concentration), mass/volume
T = absolute temperature
R = gas constant
ΔHv = solvent enthalpy of vaporization
ΔHv = solvent enthalpy of fusion
ΔTb = boiling-point elevation
ΔTf = freezing-point depression
π = osmotic pressure
ρ = density
It is important to note that the thermodynamic equations apply only for ideal solutions, a condition that can be reached only in the limit of infinite dilution of the solute. For real situations, a dilute solution can be assumed in most cases if the concentration, c, is less than 1 wt%. It should also be noted that for these techniques to work, the polymer must be fully dissolved in a good solvent, whereas polymers precipitate in a bad solvent. If not dilute enough, polymer chains may interact in the solution (Figure 5.1), causing serious error in molecular weight estimation. In each of these equations, data collected for different concentrations can be plotted versus concentration, with the value at c = 0 used to determine .
Freezing-point depression and boiling-point elevation require precise measurements of very small temperature differences. Although they are used occasionally, the difficulties involved have prevented their widespread application. For example, a Δ Tf of 0.0002 °C might be the only observed change for a polymer with molecular weight 25,000. Osmotic pressure, on the other hand, is the most common absolute method of determining . A schematic diagram of an osmometer is shown in Figure 5.2. The solution and solvent chambers are separated by a “semipermeable” membrane, one that ideally allows passage of solvent molecules but not solute molecules. The solvent flows through the membrane to dilute the solution. This is a natural consequence of the tendency of the system to increase its entropy, which is accomplished by the dilution of the solution. This dilution continues until the tendency toward further dilution is counterbalanced by the increased pressure in the solution chamber. At this point, the chemical potential of the solvent is the same in both chambers and the pressure difference between the chambers is the osmotic pressure, π. By making measurements at several concentrations, plotting (π/c) versus c and extrapolating to zero concentration, the number-average molecular weight is established through Equation (5.11).
One of the major difficulties with membrane osmometry is finding suitable semipermeable membranes. Ordinary cellophane or modifications of it are commonly used. Unfortunately, if the membrane is sufficiently tight to prevent the passage of low molecular weight chains, the rate of solvent passage is slow and it takes longer to reach equilibrium. In practice, all membranes allow some of the low molecular weight polymer in a distribution to sneak through. Also, as the average molecular weight increases, π increases, so the measurement precision also decreases. These factors usually limit the applicability of osmometry to 50,000 < < 1,000,000.
High-speed, automated membrane osmometers monitor the fluid volume in one of the chambers and externally apply the osmotic pressure to the solution chamber to prevent flow. Since no flow through the membrane is necessary, they can reduce measurement time from days to hours or even minutes.
A related colligative technique is vapor-pressure osmometry. Two thermistors are placed in a carefully thermostatted chamber that contains a pure-solvent reservoir so that the atmosphere is saturated with solvent vapor. A drop of solvent is placed on one thermistor and a drop of polymer solution is placed on the other. Because of the solvent's lower chemical potential in the solution, solvent vapor condenses on the solution drop, giving up its heat of condensation, warming the solution drop relative to the pure solvent drop. In principle, the equilibrium ΔT is thermodynamically related to the molar solution concentration, thereby allowing calculation of . In practice, heat losses (mainly along the thermistor leads) require that the instrument be calibrated for precise results, really making it a relative technique (as opposed to absolute). On a routine basis, commercial instruments are probably limited to maximum values of 40,000–50,000 [4].
5.3.1.3 Weight-Average Techniques
The absolute techniques discussed to this point establish the number of molecules present per unit volume of solution, regardless of their size or shape. Other methods measure quantities that are related to the average mass of the molecules in solution, thereby giving the weight-average molecular weight. One of the more common of these is light scattering, which is based on the fact that the intensity of light scattered by a polymer molecule is proportional, among other things, to the square of its mass. A light-scattering photometer measures the intensity of scattered light as a function of the scattering angle (Figure 5.3). Measurements are made at several concentrations. By a double extrapolation to zero angle and zero concentration (known as Zimm plot) and with a knowledge of the dependence of the solution refractive index versus concentration, is established. This technique also provides information on the solvent–polymer interaction and on the configuration of the polymer molecules in solution, since the quantitative nature of the scattering depends also on the size of the particles. Light scattering is generally applicable over a wider range 10,000 < < 10,000,000. The weight-average molecular weight can also be obtained with an ultracentrifuge, which distributes the molecules according to their mass in a centrifugal force field.
It must be noted that for light-scattering and sedimentation (by ultracentrifuge) techniques (as well as most relative methods, below) to give an accurate measurement of molecular weight, dilute solutions are required. If two polymer molecules are entangled, they will appear as a single molecule and skew the results to higher molecular weights (that is why many techniques extrapolate to zero concentration).
5.3.2 Relative Methods
The molecular weight determination techniques discussed to this point allow direct calculation of the average molecular weight from experimentally measured quantities through known theoretical relations. Sometimes, however, these relations are not known, and although something is measured that is known to be related to molecular weight, one of the absolute methods above must be used to calibrate the technique.
A case in point is solution viscosity. It has long been known that relatively small amounts of dissolved polymer could cause tremendous increases in viscosity (ever made Jell-O®?). It is logical to assume that, other things being equal, larger molecules will impede flow more than smaller ones and give a higher solution viscosity. The solution viscosity will also depend on the solvent viscosity, temperature, solute concentration, as well as the particular polymer and solvent, because the interactions between the polymer and the solvent influence the conformation of the polymer molecules in solution and entanglements between the polymer molecules. In functional form, we may write
We can get rid of the effect of solvent viscosity by calculating the fractional increase in viscosity caused by the added solute (polymer) through the specific viscosity ηsp:
(5.12)
where ηr = η/ηs is known as the relative viscosity. Similarly, we can normalize for concentration by dividing the specific viscosity by concentration to get the reduced viscosity, ηred:
(5.13)
To get rid of the influence of entanglements on viscosity, we extrapolate the reduced viscosity to zero concentration to get the intrinsic viscosity [η]:
The intrinsic viscosity, then, should be a function of the molecular weight of the polymer in solution, the polymer–solvent system, and the temperature. If measurements are made at constant temperature using a specified solvent for a particular polymer, it should be quantitatively related to the polymer's molecular weight.
Huggins proposed a relation between reduced viscosity and concentration for dilute polymer solutions (ηr < 2):
Interesting enough, k' turns out to be approximately equal to 0.4 for a variety of polymer–solvent systems, providing a convenient means of estimating dilute solution viscosity versus concentration if the intrinsic viscosity is known. By expanding the natural logarithm in the definition of inherent viscosity, ηinh, into a power series it may be shown that an equivalent form of the Huggins equation is
where
Equation From (5.14) it is seen that an alternative definition of intrinsic viscosity is
Plots of the reduced and inherent viscosities are linear with concentration, at least at low concentrations, in accord with Equations (5.16) and (5.17), and have a common intercept, the intrinsic viscosity (Figure 5.4). Exceptions occur with polyelectrolytes, where the degree of ionization and, therefore, the chemical nature of the polymer changes with concentration.
Note that the intrinsic viscosity has dimensions of reciprocal concentration. For some strange reason, concentrations were traditionally given in g/dL (=100 mL), although g/mL is now becoming more common. In fact, the relative, specific, reduced, intrinsic, and inherent viscosities are not true viscosities, and do not have dimensions of viscosity. More appropriate terminology has been proposed, but not widely adopted. Table 5.2 summarizes the various quantities defined and typical units.
Now that the intrinsic viscosity has been established, how is it related to molecular weight? Studies of the intrinsic viscosity of essentially monodisperse polymer fractions whose molecular weights have been established by one of the absolute methods indicate a rather simple relation (Figure 5.5):
where the subscript x refers to a monodisperse sample of a particular molecular weight. Equation (5.18) is known as the Mark–Houwink–Sakurada (MHS) relation.
What about an unfractionated, polydisperse sample? Experimentally, the measured intrinsic viscosity of a mixture of monodisperse fractions is a weight average
(5.19)
A viscosity-average molecular weight, , is defined in terms of this measured intrinsic viscosity as
With 0.5 < a < 1.0 for polymers, < < , but is closer to than . If the molecular weight distribution of a series of samples does not differ too much, that is, if the ratios of the various averages remain nearly the same, approximate equations of the type [η] = K′()a′ may be applicable.
Why bother with calculating if it is different from either of the molecular weight averages that we defined (Eqs (5.3) and (5.4a) for and )? Viscosity is an easily measured property of polymer solutions, and can be used for quality control in polymerization processes, so it is still quite useful.
ηr | c (g/100 mL) |
1.170 | 0.275 |
1.215 | 0.344 |
1.629 | 0.896 |
1.892 | 1.199 |
The regression slope of the upper line is 0.137. From Equation (5.17), k′ = slope/[η]2 = 0.408 (dimensionless).
Viscosities for molecular weight determination are usually measured in glass capillary viscometers, in which the solution flows through a capillary under its own head. Two common types, the Ostwald and Ubbelohde, are sketched in Figure 5.6. (Since polymer solutions are non-Newtonian, intrinsic viscosity must be defined, strictly speaking, in terms of the zero-shear or lower-Newtonian viscosity (see Chapter 14). This is rarely a problem, because the low shear rates in the usual glassware viscometers give just that. Occasionally, however, extrapolation to zero-shear conditions is required.)
In these types of viscometers, fluid is drawn up through a capillary into bulbs, with timing lines marked on the glass. As the fluid flows back down through the capillary (the right side of both viscometers in Figure 5.6), the flow time is recorded. These flow times are related to the viscosity by an equation of the form
(5.21)
where a and b are instrument constants (found through calibration), ρ is the solution density, and ν is the kinematic viscosity (with typical units of cm2/s). The last term (b/t), the kinetic energy correction, is generally negligible for flow times of over a minute, and since the densities of the dilute polymer solutions differ little from that of the solvent, the ratio of the viscosity of the polymer solution to the solvent viscosity can be approximated by the flow times recorded for these solutions, respectively:
(5.22)
The Ubbelohde viscometer has the distinct advantage that the driving fluid head is independent of the amount of solution in it; hence, dilution can be carried out right in the instrument.
The equipment necessary for intrinsic viscosity determination is inexpensive, and the measurements straightforward and rapid. The MHS constants K and a in Equations (5.18) and (5.20) are extensively tabulated for a wide variety of polymer–solvent systems and temperatures (Table 5.3) [6]. Even more simple devices (such as cup viscometers–which are metal cups with fixed holes in the bottom) are commonly used in industry for rapid viscosity estimation, this is particularly true in paint formulations. Automated viscometers, such as rotational viscometers, are also commonly used in analytical laboratories and offer rapid and accurate results.
5.4 Molecular Weight Distributions
Before continuing to one additional molecular weight characterization technique (gel permeation chromatography), let us delve a bit deeper into the concept of molecular weight distributions. A typical synthetic polymer might consist of a mixture of molecules with degrees of polymerization x ranging from one to perhaps millions. The complete molecular weight distribution specifies the mole (number) or mass (weight) fraction of molecules at each size level in a sample. (Actually, moles or masses could be specified, but they vary linearly with sample size, that is, are extensive quantities, while the mole or mass fractions are intensive, independent of sample size, and are therefore preferable.)
Distributions are often presented in the form of a plot of mole (nx/N) or mass (wx/W) fraction of x-mer versus either chain length, x, or chain mass, Mx. Since x and Mx differ by a constant factor, the molecular weight of the repeating unit Mr (Eq.(5.1)), it makes little difference which is used. Because x can assume only integral values, a true distribution must consist of a series of spikes, one at each integral value of x, or separated by Mr molecular weight units if plotted against Mx. The height of the spike represents the mole or mass fraction of that particular x-mer. No analytical technique is capable of resolving the individual x-mers, so distributions are drawn (and represented mathematically) as continuous curves, drawn through the tops of the spikes as sketched in Figure 5.7.
The averages may be calculated from either the mole- or the mass-fraction distributions. For continuous distributions, the summations in Equations (5.3) and (5.4) must be replaced by the corresponding integrals (the subscript x has been dropped for clarity and because of the continuous nature of the distributions).
From the mole (number)-fraction distribution (n/N),
Since = m and M = mx:
Note that the integrals replace the summation symbols used in the definition of (Eq.(5.3)). Also,
(5.25)
that is, the mole fractions must sum to 1. Also,
And, since = m,
(5.27)
If the mass (weight)-fraction distribution (w/W) is known, the average molecular weights and chain lengths can be determined:
(5.28)
(5.29)
(5.30)
(5.31)
Note that
(5.32)
that is, the mass fractions must sum to 1.
Given the mole (number)-fraction distribution, the mass (weight)-fraction distribution may be calculated, and vice versa. Since w = nM and, by definition of (Eq.(5.3)), W = N,
5.5 Gel Permeation (Or Size-Exclusion) Chromatography (GPC, SEC) [7]
Until fairly recently, the approximate molecular weight distribution of a polymer could be determined only by laborious fractionation of the sample according to molecular weight followed by determination of the molecular weights of the individual fractions with one of the techniques previously discussed. The bases for these fractionation techniques are discussed in Chapter 7. Suffice it to say that they tend to be difficult and time-consuming and so, in general, are avoided whenever possible.
Now, however, gel permeation chromatography (GPC, also known as size-exclusion chromatography, SEC) has been firmly established as a means of rapidly determining molecular weight averages (, , ) and the corresponding distributions. GPC makes use of a column, or series of columns, packed with particles of a porous substrate (known as the stationary phase). The term gel in gel-permeation chromatography refers to a crosslinked polymer that is swollen by the solvent used. This is perhaps the most common type of substrate, but others, for example, porous glass beads, are also used. The column is maintained at a constant temperature, and the solvent (known as the mobile phase) is passed through it at a constant rate. At the start of a run, a small amount of polymer solution is injected into the flowing mobile phase just before it enters the column. The solvent flow carries the polymer through the column. The smaller molecules in the sample have easy access to the substrate pores and diffuse in and out of the pores, creating a highly tortuous, lengthy pathway from the entrance to the exit from the column. The larger polymer chains simply cannot fit into the pores in the stationary phase and pass around the outside, and because they have a shorter pathway from entrance to exit, they elute from the column first. Thus, polymer chains of different molecular weights are effectively separated, with the molecular weight of the chains exiting the column decreasing with time.
A concentration-sensitive detector is placed at the outlet of the column. The most common detector is a differential refractometer (which measures the difference in refractive index between the pure solvent and that of the eluting fluid, containing the polymer), but ultraviolet (UV) or infrared (IR) detectors are also useful as long as the polymer has a functional group that can be detected in these light ranges. Regardless of the type of detector, it is essential that it measure some quantity Q that is proportional to the mass concentration of polymer at the column outlet (Q must be independent of M):
Thus, a GPC curve consists of a plot of Q (usually in arbitrary scale divisions, but related to the quantity of chains eluting at a particular time) versus v, the volume of solvent that has passed through the detector since sample injection, called elution or retention volume. The x-axis can easily be converted to time by dividing the elution volume by the volumetric flow rate of the solvent.
An example of a GPC curve is shown in Figure 5.9. Such a curve is often immediately useful for quality-control purposes, as the width of the distribution indicates sample polydispersity. The graph can also be compared directly to the graph for a known standard material, revealing a quick qualitative comparison. This may be enough to take corrective action in processing, but the usefulness of GPC extends well beyond qualitative comparisons, because, with appropriate calibration, it can determine all types of average molecular weights (, , ), and the PI of polymer samples. An additional benefit is that GPC can also be useful in detecting small molecular weight impurities or additives (such as plasticizers or stabilizers––see Chapter 18). These materials show up as peaks at the low molecular weight (high v or long t) end of the spectrum.
GPC chromatograms might look like one of the curves plotted in Figure 5.9. Nearly monodisperse polystyrene standards are used to calibrate a GPC. If run through the column as a mixture, the elution curve will show sharp peaks for each molecular weight standard, with the highest molecular weight eluting first. After the column has been calibrated, polymers with unknown molecular weight distributions can be tested. In Figure 5.9, one sample with two narrowly dispersed samples are shown in curve B (note that the curves do not overlap and that the area under the curves is proportional to the amount of polymer of each molecular weight), while a mixture of highly polydisperse polymer samples will have very broad peaks that overlap (curve C).
The total area under a chromatographic peak is proportional to the amount (mass) of polymer eluted. Thus, for polymer mixtures, it is possible to use GPC to estimate the weight fractions of different components, as long as their peaks do not overlap on the elution (x) axis.
It should be emphasized that GPC requires dilute solutions, so that each polymer molecule passes through the porous stationary phase with no entanglements. Because the hydrodynamic volume occupied by grafted or branched polymers is somewhat less than that occupied by a linear polymer with the same molecular weight, GPC typically underpredicts and for nonlinear polymers.
Conventional GPC is a relative method. Thus, to provide quantitative results, the relation between M and v (or elution time, t) must be established by calibration with monodisperse polymer standards. A calibration curve is shown in Figure 5.10. Typically, when the data are plotted in the form log M versus v, the curve is linear over much of the range, but sometimes turns up sharply at low v (high M).
The calibration curve, strictly speaking, applies only to the particular polymer, solvent, temperature, flow rate, and column for which it was established. Change any one, and the calibration is no longer valid. Most SEC calibrations are obtained with polystyrene, because the necessary monodisperse standards are readily available at reasonable cost. What do you do when you want to analyze another polymer? If absolute values are not needed, the molecular weights can be reported with respect to the polystryene (or other polymer) standards used to calibrate. A more sophisticated approach was suggested by Grubisic et al. [8]. When they plotted log([η]M) (the product [η]M is proportional to the hydrodynamic volume of the molecules) versus v, they obtained a single, universal calibration curve for a variety of different polymers. Thus, if you measure intrinsic viscosity along with SEC elution volume for your polystyrene standards (or calculate [η] for them with Eq. (5.18) and literature values for K and a), you can plot the universal calibration curve for your column.
To back out a calibration curve for a different polymer, you would need to know K and a under the new conditions:
(5.35)
(5.36)
where the subscript 0 above refers to the calibration conditions, that is, values from the universal calibration.
Balke et al. [9] have proposed a calibration method that requires only a single polydisperse sample of known and . The method assumes that the relation between log M and v is linear, and therefore it can be characterized by two parameters, a slope and an intercept. Given the GPC curve and the calibration, and can be calculated (using techniques outlined below). With the GPC data on the standard, a computer program adjusts the two calibration parameters until the known and are obtained, in effect using the two known average molecular weights to solve for the two unknown calibration parameters. If the calibration curve is not linear (and you would have no way of knowing), serious errors can result.
Most GPC packages include the software necessary to calculate molecular weight averages and the complete distribution from the data and the calibration, so one needs only to push the appropriate buttons. Nevertheless, the equations are outlined below to provide an understanding of what the software is doing and what its limitations might be.
Molecular weight averages may be approximated directly from the GPC curve and calibration by breaking the curve into arbitrary volume increments Δv. Usually, Δv is taken as 5 cm3, but smaller Δv's improve accuracy. The number of moles of polymer in a volume increment Δv is ni:
(5.37)
where ci = polymer mass concentration in the ith volume increment
Mi = molecular weight of polymer in the ith volume increment (assumed essentially constant over small Δv)Because ci = Qi/k (Eq.(5.34)),
If the volume increment Δv is constant, inserting Equation (5.38) into Equations (5.3) and (5.4) gives
Note that the proportionality constant k between concentration and GPC readout cancels out, and therefore need not be known. Qi is normally read from the GPC curve as the distance above the baseline (in any convenient units) at a given v. The baseline presumably represents the detector output with the pure solvent. Establishing a good baseline is not always a trivial procedure, and the results can be quite sensitive to its location. Mi is read from the calibration curve at the same v. For greater accuracy, if necessary, the integral analogs of Equations (5.39) and (5.40) may be used, with the integrals evaluated numerically:
In most cases, the averages as calculated above and the qualitative information about the distribution provided by the GPC curve are all that are necessary. However, the technique is capable of providing the true distributions, if desired. The necessary calculations are outlined below:
(5.43)
(5.44)
(5.45)
Note:
The preceding calculations assign all the moles of polymer in a range m of the continuous GPC curve to a single spike (see Section 4) to calculate the correct height of the distribution:
(5.46)
(5.48)
(5.49)
(5.50)
Again, k cancels out in these equations. Once the distributions are known, the exact averages may be calculated by the techniques in Section 4. For example, insertion of Equation (5.47) into Equations (5.23) and (5.26) gives Equations (5.41) and (5.42).
GPC columns can be selected that have a wide spectrum of pore sizes (and thus separate a wide range of molecular weights–from thousands to millions). Alternatively, columns with more narrow pore size distributions can be used to get a better separation of polymers that have similar molecular weights—say from 1000 to 20,000. These columns are designed to be used with organic solvents (required for most polymers) or water that flow through as the mobile phase. The selection of which columns to use depends on the polymer structure and the expected molecular weight distribution.
A multiangle laser light-scattering photometer is often added to the usual concentration detector. This combination performs an online light-scattering determination of the M of the polymer in the eluting stream. This makes GPC an absolute technique, eliminating the need for calibration. Similarly, a differential viscometer is commercially available which performs an online intrinsic viscosity measurement on the column effluent [10]. Recall that the ordinate of the universal calibration is [η]M. This device requires simple calibration (in effect locating the universal calibration curve for the particular column, temperature, solvent, etc.) to provide absolute molecular weight distributions and also provides information on the hydrodynamic volume of the molecules in solution and therefore on branching, because a branched molecule in solution has a smaller hydrodynamic volume than a linear molecule of the same molecular weight. As mentioned earlier, an important caution regarding GPC/SEC is as follows: because the molecules are separated based on their size, linear polymers will flow differently through the stationary phase (the packed column) than branched or grafted polymers, so molecular weight estimation by this technique is normally less accurate for branched, grafted, or dendritic (very highly branched) polymers.
5.6 Summary
Molecular weight has an enormous impact on polymer properties, as we will discover in later chapters. Because polymers are made of many copies of the same repeat unit, traditional methods to determine absolute molecular weights are not applicable. Thus, several techniques are discussed in this chapter to measure or estimate polymer molecular weights. Depending on the techniques used, the number-average or weight-average molecular weights are determined, with most sensitive techniques requiring that the polymer be in dilute solution so that physically entangled polymers are not mistakenly recorded as one large polymer. Of the techniques discussed so far, gel permeation chromatography offers the greatest amount of information about a polymer sample, giving the size distribution of a polymer sample, which allows calculation of and and the polydispersity index.
Problems
Cussler | Split end | 180 lb |
Miller | Left tackle | 270 lb |
Westerberg | Left guard | 256 lb |
Brenner | Center | 285 lb |
Anderson | Right guard | 260 lb |
Jain | Right tackle | 305 lb |
Prieve | Tight end | 250 lb |
Concentration, CA g/mL | Osmotic Pressure, πA, cm of Solvent |
0.0320 | 0.700 |
0.0660 | 1.82 |
0.100 | 3.10 |
0.140 | 5.44 |
0.190 | 9.30 |
1. Martin, J.R., J.F. Johnson, and A.R. Cooper, J. Macromol. Sci. Rev. C8(1), 57 (1972).
2. Nunes, R.W., J.R. Martin, and J.F. Johnson, Polym. Eng. Sci. 22(4), 193 (1982).
3. Collins, E.A., et al., Experiments in Polymer Science, Wiley, New York, 1973.
4. Burge, D.E., American Laboratory Raw data from GPC Chromatogram of Polystyrene 9(7), 41 (1977).
5. Billmeyer, F., Textbook of Polymer Science, 2nd ed., Wiley, New York, 1971, Chapter 3.
6. Kurata, M. and Y. Tsunashima, Viscosity-molecular weight relationships and unperturbed dimensions of linear chain molecules, Chapter VII/1 in Polymer Handbook, 3rd ed., J Brandrupand E.H. Immergut (eds), Wiley, New York, 1975.
7. Provder, T. (ed.), Size Exclusion Chromatography: Methodology and Characterization of Polymers and Related Materials, ACS Symposium Series No.245, American Chemical Society, Washington, DC, 1984.
8. Grubisic, Z. et al., J. Polym. Sci. B-5, 753 (1967).
9. Balke, S.T., et al., Ind. Eng. Chem., Prod. Res. Dev. 8, 54 (1969).
10. Yau, W.W. and S.W. Rementer, J. Liquid Chromatogr. 13(4), 626; Wang, P.J. and B.S Glassbrenner, ibid, 11(16) 3321 (1988).